Manually Setting the IP Address
When connected to a network, the MZA 6010 screen will obtain its own IP Address. It does this during the power-up procedure, so if communication via Ethernet is preferred, plug the cable into the instrument before turning it on. To use a specific IP address, it must be entered on the Windows CE screen. This can only be accessed when the Gas Analyzer software has been shut down.
-
1. Log in to the MZA 6010 touch screen using the supervisor access code (by default, this code is ‘2’).
-
2. Access the Instrument Setup → General Setup menu.
-
3. Select “Shut Down Software” followed by “Yes”.
-
4. The program will shut down and a Windows CE screen will appear.
-
5. Click the Start menu in the lower left corner.
-
6. Select ‘Settings” and then “Network and Dial-up Connections”.
-
7. Select “CS89001”
-
8. Select the option for “Specify an IP Address”
-
9. Enter the desired IP Address, Subnet Mask, and Default Gateway.
-
10. Select “OK”
-
11. Close open windows to return to the main Windows CE Screen.
-
12. Double-click on the “SaveRegs” icon.
-
13. Select “OK” from the Registry Save Complete notification.
-
14. Turn the instrument off and wait for the screen to turn off.
-
15. Turn the instrument back on.
-
16. The instrument will automatically boot up as a gas analyzer with the IP address that has been manually entered.